The evaluation of system accuracy by lDK revealed that meisurements obtained with the
blood glucose meter fulfill the acceptance criteria asspecified by DIN EN ISO 15197:2015.
Combined Bland-Altman Diagram
Consensus Error Grid
Product description | BG-709 Blood Glucose Monitoring System | Size | 90*61*23.6mm(L*W*H) |
Assay method | Glucose dehydrogenase biosensor | Sheif life | Three years |
Measurement range | 0.5-33.3mmol/L(9~600mg/dL) | LCD size | 35.4*30.9mm(L*H) |
Sample | Fresh capillary whole blood, venous blood, neonatal blood, arterial blood | Weight | Approximate 51.6g, battery not included |
Sample size | Approximate 0.6 microlitre | Operating environment | 5℃~45℃(41F~113F)10 ~ 90% RH(non-condensing) |
Response time | 5 seconds | ||
Calibration | Plasma equivalent glucose values | Meter storage enironment | -20℃~55℃(-4F~131F)10 ~ 95% RH(non-condensing) |
Battery | 2xAAA alkaline batteries | ||
Battery life | Approximately 1000 tests | Altitude | Up to 10000 feet(3048 meters)above sea level |
Unit of measure | mmol/L or mg/dL; switchable | Automatic power off | Approx. 3 minutes after last user action |
Memory | 360 blood glucose results with date and time | Strip storage environment | 1℃ ~ 30℃ (33.8F ~ 86F) |
Hematocrit | 0% ~ 70% |
① lnsert Test Strip into Meter
② Blood Sampling
③ Apply Blood Sample
④ Read Results in 5 Seconds