NEED HELP?
Write your message here and send it to us
The evaluation of system accuracy by IDK revealed that meisurements obtained with the
blood glucose meter fulfill the acceptance criteria asspecified by DIN EN ISO 15197:2015.
Combined Bland-Altman Diagram
Consensus Error Grid
Product description | BG-715b Eve Blood Glucose Monitoring System | Size | 94*60*21mm (L*W*H) |
Assay method | Glucose dehydrogenase biosensor | Sheif life | Three years |
Measurement range | 0.5-33.3mmol/L(9~600mg/dL) | LCD size | 40*40mm (L*H) |
Sample | Fresh capillary whole blood, venous blood, neonatal blood | Weight | Approximate 62 g, battery not included |
Sample size | Approximate 0.6 microlitre | Operating environment | 5℃~45℃(41F~113F)10 ~ 90% RH(non-condensing) |
Response time | 5 seconds | ||
Calibration | Plasma equivalent glucose values | Meter storage enironment | -20℃~55℃(-4F~131F)10 ~ 95% RH(non-condensing) |
Battery | 2xAAA alkaline batteries | ||
Battery life | Approximately 1000 tests | Altitude | Up to 10000 feet(3048 meters)above sea level |
Unit of measure | mmol/L or mg/dL; switchable | Strip storage environment | 1℃ ~ 30℃ (33.8F ~ 86F) |
Memory | 500 blood glucose results with date and time | Hematocrit | 0%~70% |
① lnsert Test Strip into Meter
② Blood Sampling
③ Apply Blood Sample
④ Read Results in 5 Seconds