NEED HELP?
Write your message here and send it to us
The evaluation of system accuracy by lDK revealed that meisurements obtained with the
blood glucose meter fulfill the acceptance criteria asspecified by DIN EN ISO 15197:2015.
Reference value [laboratory analyzer], Hemoglobin Concentration (g/dL)
Market Leader Chemistry Analyzer (g/L)
Meter Type | HB-101 | Meter storage environment | -20℃~55℃(-4F~131F)
<90%RH(non-condensing) |
Blood Sample | Capillary and Venous whole blood | ||
Blood Volume | 10μL | Strip storage environment | 2℃ ~ 30℃ (35.6F ~ 86F) |
Test Strip Type | HS-101 | Dimensions | 126*57*24mm |
Measuring Range | 4.5-25.6g/dL,45-256g/L,2.8-15.9mmol/L | Repeatability | 4.5-10.0g/dL,Standard Deviation(SD)<=0.8g/dL
10.1-25.6g/dL,Coefficient of Variation(CV)<=3% |
Response time | <10 seconds | ||
Power | 3*AAA Battery or MiniUSB DC5V | Accuracy | 4.5-10.0g/dL,absolute deviation<=1g/dL
10.1-25.6g/dL,relative deviation <= 4% |
Battery Life | Approximately 2000 tests | ||
Memory Capacity | 1000 test results | Weight | 185g(with battery) |
Operation Environment | 10℃ ~ 40℃(50F~104F)
<90%RH(non-condensing) |
Expected service life | Ten years |
Automatic power off | Approx.3 minutes after last user action | ||
Unit of measure | g/dL,g/L or mmol/L |