The evaluation of system accuracy by lDK revealed that meisurements obtained with the
blood glucose meter fulfill the acceptance criteria asspecified by DIN EN ISO 15197:2015.
Reference value [laboratory analyzer], Hemoglobin Concentration (g/dL)
Market Leader Chemistry Analyzer (g/L)
Meter Type | HB-201/202 | Meter storage environment | -20℃~55℃(-4F~131F)<90%RH(non-condensing) |
Blood Sample | Capillary and Venous whole blood | ||
Blood Volume | 10μL | Strip storage environment | 2℃ ~ 30℃ (35.6F ~ 86F) |
Test Strip Type | HS-101 | Dimensions | 126*71*25mm |
Measuring Range | 4.5-25.6g/dL,45-256g/L,2.8-15.9mmol/L | Repeatability | 4.5-10.0g/dL,Standard Deviation(SD)<=0.8g/dL10.1-25.6g/dL,Coefficient of Variation(CV)<=3% |
Response time | <10 seconds | ||
Power | 3*AAA Battery or MiniUSB DC5V | Accuracy | 4.5-10.0g/dL,absolute deviation<=1g/dL10.1-25.6g/dL,relative deviation <= 4% |
Battery Life | Approximately 2000 tests | ||
Memory Capacity | 1000 test results | Weight | 113.6g(battery not included) |
Operation Environment | 10℃ ~ 40℃(50F~104F)<90%RH(non-condensing) | Expected service life | Five years |
Automatic power off | Approx.3 minutes after last user action | ||
Unit of measure | g/dL,g/L or mmol/L |